Skip to main content
Laboratory & Diagnostics11 min read

Laboratory Quality Control: A Practical Westgard Rules Guide

A working guide to laboratory quality control: setting the mean and SD, reading Levey-Jennings charts, choosing Westgard multirules by test precision, run rejection policy, and handling an unacceptable EQA result.

Dr. Shalini Deshmukh

Hospital Accreditation and Quality Consultant

#westgard rules#levey jennings chart#internal quality control#laboratory qc#proficiency testing
Laboratory Quality Control: A Practical Westgard Rules Guide

What laboratory quality control is actually asking

Internal laboratory quality control asks a narrow question: on this run, on this instrument, is the analytical process behaving the way it behaved when we characterised it? It does not ask whether the result is clinically correct, and it does not detect errors that occurred before the sample reached the analyser. Understanding that boundary prevents the two opposite mistakes of trusting QC too much and dismissing it as ritual.

A control material with an established mean and standard deviation is run alongside patient samples. The observed control value is compared against decision rules. If the rules are violated, the run is held, patient results are not released, the cause is investigated, and the run is repeated after correction. That is the entire mechanism, and everything else is detail about which rules to apply and when.

The reason the detail matters is that rules trade sensitivity against false rejection. A rule set that catches every real shift will also stop good runs and exhaust the staff who have to investigate them. A rule set that never triggers is not protecting anyone. Choosing well is a quality decision, not a technical formality.

Medical technologist reviewing internal quality control results before releasing a batch of patient reports
Medical technologist reviewing internal quality control results before releasing a batch of patient reports

The mean and SD you start with determine everything after

Control limits are only as meaningful as the mean and standard deviation behind them. Manufacturer-assigned values on the control insert are a starting point for a new lot, not a permanent target — they are derived across many instruments and methods and are typically wider than your own performance. Using them indefinitely produces charts that almost never flag anything.

The laboratory should establish its own mean and SD from a run of its own data on the specific instrument and method, collected across multiple days so that between-day variation is represented. Twenty independent points is a common minimum, and the values should be reviewed when reagent lot, calibrator lot, or instrument service changes materially.

Lot changeover deserves its own routine. Overlapping the new control lot with the old one for a period lets you establish the new mean against known-stable performance instead of assuming a shift was caused by the lot. Skipping the overlap is how laboratories end up unable to distinguish a control problem from an analytical one.

Reading a Levey-Jennings chart without over-reacting

A Levey-Jennings chart plots each control value against the mean with lines at one, two, and three standard deviations. Its usefulness is pattern recognition over time, not the judgement of any single point. A run of values drifting steadily upward tells you something quite different from a single point beyond two SD, even though the single point looks more alarming.

The two patterns worth naming are shift and trend. A shift is an abrupt move to a new level that then holds — typically a new reagent lot, a recalibration, or a maintenance event. A trend is a gradual movement across successive runs — typically deteriorating reagent, a drifting light source, an ageing electrode, or a slowly failing temperature control.

Charts are also how you notice the least dramatic failure mode: control values that hug the mean too tightly. Unnaturally low scatter usually means the control is not being treated like a patient sample, or that someone is repeating until the number looks acceptable. That behaviour destroys the entire purpose of the exercise.

Levey-Jennings control chart showing a gradual upward trend across successive analytical runs
Levey-Jennings control chart showing a gradual upward trend across successive analytical runs

Choosing Westgard multirules by test precision

Westgard multirules exist because a single rule cannot balance error detection against false rejection across tests with different analytical performance. Applying one rule set to every analyte is the most common misconfiguration in hospital laboratories. A test with wide allowable error and tight precision needs far less rule machinery than a test running close to its clinical limit.

The practical approach is to compare each test's imprecision against the total allowable error for that analyte, then select rules accordingly. Where the method is comfortably better than the requirement, a simple rule with two control levels is sufficient and will rarely produce false alarms. Where the method is marginal, multirules with more control observations are justified because the process genuinely needs closer watching.

Document the selection per analyte with the reasoning attached. Assessors ask why a rule set was chosen, and the answer should be a performance comparison rather than a vendor default. This also gives the laboratory a defensible basis for relaxing rules on tests that keep generating investigations without ever finding a real cause.

Common multirules and what each detects

  • 1-3s: a single point beyond three SD, sensitive to sudden random error
  • 2-2s: two consecutive points beyond the same two-SD limit, indicating systematic shift
  • R-4s: two points in a run spanning four SD, indicating increased random error
  • 4-1s: four consecutive points beyond the same one-SD limit, an early systematic signal
  • 10x: ten consecutive points on the same side of the mean, a persistent bias
  • 1-2s used as a warning rather than a rejection, triggering inspection of the other rules

Run rejection, repeat policy, and result release

A rejection rule has meaning only if the laboratory has decided in advance what happens next. The policy should state that patient results from the affected run are held, that the cause is investigated before repeating, and that a repeat is permitted only after a defined corrective action. Repeating the control first and thinking later is the habit that turns quality control into theatre.

Where the investigation identifies a cause — expired reagent, a missed maintenance step, a calibration due — the corrective action is obvious and the run is repeated. Where no cause is found, the laboratory needs a stated limit on how many repeats are allowed before escalation. Two is a common ceiling, and escalation means a section head decision recorded with the run.

Held results have clinical consequences, so the policy needs a communication route. Where a run affects urgent or critical care samples, the treating team should be told that results are delayed rather than left waiting. A laboratory system such as HealUDoc can hold results in an unreleased state and surface the delay to the ordering clinician, which is preferable to a silent gap in the record.

External quality assessment and proficiency testing

Internal control tells you whether today matches your own history. External quality assessment tells you whether your history is correct. Both are required, and neither substitutes for the other — a laboratory can be beautifully precise around a wrong mean, and only an external comparison will reveal it.

Participation should cover the laboratory's scope, including tests with low volume, because rarely performed tests carry higher risk rather than lower. Where no scheme exists for a particular analyte, interlaboratory comparison with a peer laboratory using a documented protocol is the accepted alternative, and it needs the same record trail as formal participation.

The most important operational rule is that PT samples must be handled exactly like patient samples — same staff, same instrument, same procedure, no repeat testing, no consultation with other laboratories before submission. Special handling of PT material invalidates the whole exercise and is treated seriously by accreditation bodies.

Laboratory team reviewing an external quality assessment report against internal control performance
Laboratory team reviewing an external quality assessment report against internal control performance

Handling an unacceptable EQA result

An unacceptable external result is a documented non-conformity and should be treated with the same investigation discipline as a clinical incident. Start by confirming that it is real: check for transcription errors, unit mismatches, sample mix-up, and whether the reported value was submitted for the correct analyte. Clerical causes are common and are still causes worth recording.

If the value was genuinely produced, the investigation moves to method, calibration, reagent lot, and instrument state on the date the sample was run. Reviewing the internal QC charts around that date usually narrows the question quickly — a shift visible in-house that nobody actioned tells a different story from clean internal control alongside an external failure, which points toward calibration bias.

The closure step that laboratories skip is patient impact assessment. If a bias existed, the laboratory has to decide whether previously released results in that window need review or reissue, and record the decision either way. That assessment, more than the corrective action itself, is what an assessor will look for.

A failed proficiency sample is not a bad day. It is the one piece of evidence you get each cycle that your own numbers might be quietly wrong.

Consultant biochemist at a tertiary hospital laboratory
Share this article
Back to all articles

Keep reading

Related articles

See HealUDoc in action

From EHR to analytics, watch how one platform runs your entire hospital. Book a personalized walkthrough with our team.